Optics Characterization Tests ============================= UUID: ``F66D66DC-23E9-42EC-A635-1D8BE9832C16`` Optics characterization should produce test artifacts for: * Wavelength accuracy after pixel-to-wavelength calibration. * FWHM of narrow emission lines. * SNR at representative visible and near-infrared wavelengths. * Linearity versus input intensity and integration time. * Repeatability after repeated captures and after minor handling. * Stray light and second-order diffraction behavior. * Comparison with a reference source or commercial spectrometer. The test record should include optical configuration, slit width, grating setting, filter state, integration time, calibration version, and raw exported data path.